The PM500W fuel cell single cell test station can be used for the development and test of 500W PEM fuel cell single cell. This station is mainly improved to solve the dynamic response, long-term operation stability and other requirements, and can be used to test the performance and durability of the fuel cell single cell.
This station is capable of controlling the flow, pressure, temperature, and humidity of the anode, cathode, and cooling loops of a fuel cell stack. It simulates operational conditions and collects comprehensive data from the stack.
*Customized products, please send your requirements to info@tensortechs.com
● Anode/Cathode Flow Rate, Stoich, Temperature, Humidity and Pressure Control
● Dry/Wet Gas Switching
● Estop and Manual Nitrogen Purge
● Rapid Heating and Humidification of The Stack
● Polarization Curves and Other Functional Tests
● Cell Voltage Monitoring
● Intrinsic Safety & Alarming System
● HMI with Data Acquisition & Analysis
● 24/7 Unattended Operation
● Cell Hardware
● End Plate Heating
● 0V Operation
● Cell Reverse
● EIS
● HFR
●Dynamic Pressure
●Anode Dead-end
| Min P/kW | Max P/kW | H2/NLPM | Air/NLPM | Inlet DP/°C | Gas Temp/°C | Coolant/LPM | Current/A | Cell# | Stack Temp/°C | Gas Pressure /kPag | Cell Hardware | Endplate Heating | 0V Operation | Cell Reverse | EIS | HFR | Rapid Humid | Dynamic Pressure | Active Cooling | Anode Deadend | Cold Start | Water Balance |
| 0.025 | 0.5 | 20 | 40 | 90 | 100 | 2 | 750 | 4 | 90 | 300 | O | O | O | O | O | O | O | O |